American Academy of Sleep Medicine (AASM) scoring manual: Technical and digital specifications for recording movements

American Academy of Sleep Medicine (AASM) scoring manual: Technical and digital specifications for recording movements
EMG Sampling rate range 500 to 200 Hz
Filter settings Low frequency 10 Hz
High frequency 100 Hz
PSG recording software Histogram with capability to record leg movement events with cursor capability to jump from page to page
Recording video capacity with PSG synchronization at an accuracy of at least one video frame per second
EMG: electromyography; PSG: polysomnography.
Berry RB, Quan SF, Abreu AR, et al. AASM Manual for the Scoring of Sleep and Associated Events: Rules, Terminology and Technical Specifications, 2.6, American Academy of Sleep Medicine, Darien, IL 2020.